SOP measure cable loss
using Network Analyzer
Michael ( 丁丁丁 )
2021/02/02
I. Setting Network Analyzer
2
1
Select “Preset”, then select “OK” to reset the device
Confidential Material for Internal Use Only
2
I. Setting Network Analyzer
4
3
6
5
On “Display” select “Allocate traces” “x4” “Num of Traces” “4”
Confidential Material for Internal Use Only
3
I. Setting Network Analyzer
S11
S12
Return > Measurement > S11 > S12 > S21 > S22 as above picture
Confidential Material for Internal Use Only
4
S21
S22
I. Setting Network Analyzer
9
7
8
Select “Start” , “Stop” to choose the range of frequency. Ex: Start 2G, stop 6G
Select “point” “1601”
Confidential Material for Internal Use Only
5
I. Setting Network Analyzer
Select “Cal Kit” then confirm the Calibration Kit type as if correct. For example: 85033E
Confidential Material for Internal Use Only
6
II. Cable loss the Network Analyzer
10
11
12
Select “Calibration”, then select “Calibrate”, choose “2-Port Cal”
Confidential Material for Internal Use Only
7
II. Cable loss the Network Analyzer
13
14
Select “Reflection”, choose “port 1 Open”
Connect RF cable from Port1 to “O” of Calibration kit as picture,
then click “Open”
Confidential Material for Internal Use Only
8
II. Cable loss the Network Analyzer
Select “Port1 Short” then connect RF cable to “S” of calibration kit, select “Short”
Select “Port1 Load” then connect RF cable to “L” of calibration kit, select “Broadband”
Confidential Material for Internal Use Only
9
II. Cable loss the Network Analyzer
Connect RF cable from Port2 the connect with calibration kit the same do with
Port1
Confidential Material for Internal Use Only
10
II. Cable loss the Network Analyzer
17
15
Select “Transmission”, connect RF cable of port1 with
RF cable of port2 then choose “Thru”
16
Confidential Material for Internal Use Only
11
II. Cable loss the Network Analyzer
Select “Done” to finish cable loss the device
Confidential Material for Internal Use Only
12
III. Cable loss the RF cable
19
20
21
Connect RF cable which want to measure loss on 2 ports of device
Select “Marker 1”, “Marker 2”…to select the frequency
After choose frequency finish, look on screen to see the calibration value on location 20
Record the cable loss value for file setting
Note: The values at location
Confidential Material for Internal Use Only
19 21 should be small than “-20”
13