SOP measure cable loss
using Network Analyzer
Michael ( 丁丁丁 )
2021/02/02


I. Setting Network Analyzer

2

1

 Select “Preset”, then select “OK” to reset the device

Confidential Material for Internal Use Only

2


I. Setting Network Analyzer

4

3

6
5

 On “Display” select “Allocate traces”  “x4”  “Num of Traces”  “4”
Confidential Material for Internal Use Only


3


I. Setting Network Analyzer
S11

S12

 Return > Measurement > S11 > S12 > S21 > S22 as above picture

Confidential Material for Internal Use Only

4

S21

S22


I. Setting Network Analyzer
9
7
8

 Select “Start” , “Stop” to choose the range of frequency. Ex: Start 2G, stop 6G
 Select “point”  “1601”

Confidential Material for Internal Use Only

5



I. Setting Network Analyzer

 Select “Cal Kit” then confirm the Calibration Kit type as if correct. For example: 85033E

Confidential Material for Internal Use Only

6


II. Cable loss the Network Analyzer
10
11

12

 Select “Calibration”, then select “Calibrate”, choose “2-Port Cal”

Confidential Material for Internal Use Only

7


II. Cable loss the Network Analyzer

13

14


 Select “Reflection”,  choose “port 1 Open”
 Connect RF cable from Port1 to “O” of Calibration kit as picture,
then click “Open”

Confidential Material for Internal Use Only

8


II. Cable loss the Network Analyzer

 Select “Port1 Short” then connect RF cable to “S” of calibration kit, select “Short”
 Select “Port1 Load” then connect RF cable to “L” of calibration kit, select “Broadband”

Confidential Material for Internal Use Only

9


II. Cable loss the Network Analyzer

 Connect RF cable from Port2 the connect with calibration kit the same do with
Port1

Confidential Material for Internal Use Only

10


II. Cable loss the Network Analyzer


17

15

 Select “Transmission”, connect RF cable of port1 with
RF cable of port2 then choose “Thru”

16

Confidential Material for Internal Use Only

11


II. Cable loss the Network Analyzer

 Select “Done” to finish cable loss the device

Confidential Material for Internal Use Only

12


III. Cable loss the RF cable

19

20


21

 Connect RF cable which want to measure loss on 2 ports of device
 Select “Marker 1”, “Marker 2”…to select the frequency
 After choose frequency finish, look on screen to see the calibration value on location 20
 Record the cable loss value for file setting
 Note: The values at location

Confidential Material for Internal Use Only

19 21 should be small than “-20”

13



×