Agilent InfinityLab LC Series
Preventive Maintenance Checklist
Agilent InfinityLab
Preventive
Maintenance
LC Series
Checklist

Agilent Preventive Maintenance (PM) provides factory recommended service for your systems to
assure reliable operation and the accuracy of your results. Delivered by highly-trained and
certified service engineers using genuine Agilent parts and supplies, Agilent Preventive
Maintenance provides everything you need to reduce unplanned downtime and keep your systems
operating at their peak.
For more information about Agilent Technologies services please visit our web site using the
following URL />
Customer Information






Customers should provide all necessary operating supplies upon request of the engineer.
A customer representative should be available to the engineer while performing the
preventive maintenance procedures.
Any parts, not included in the Parts Lists section of this document, are not part of the
recommended Preventive Maintenance service, nor are they included in the price of this
service.
If a system requires the use of additional or special procedures and/or parts for the
instrument service, then these must be ordered separately and charged as a repair, which may
incur additional costs.



Service Engineers Responsibilities
Only complete/printout pages that relate to the system or module being serviced.
Complete empty fields with the relevant information.
Complete the relevant checkboxes in the checklist using a “X” or tick mark “ ” in the checkbox.
Complete Not Applicable check boxes to indicate services that are not delivered as needed.
Complete the PM service in the order of the task listed.
Complete the Service Review section together with the customer. In case of diagnostic tests,
record test results in section service review (test results).

Additional Instruction Notes


Please exercise care when using solvents close to the instrument. It could damage the vapor
sensor present in the ELSD.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 1 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

System Information

Guidance:
 Check box if instrument configuration report is attached instead of completing the table
below.
Instrument System Name/I.D:

Instrument Location:

Record the list of system component product numbers
below.

List the serial numbers of the components present in the
system below.

1.

1.

2.

2.

3.

3.

4.

4.

5.


5.

6.

6.

7.

7.

8.

8.

9.

9.

10.

10.

Preparation
 Consult the device user manual for further instructions.

Issued: 10-Aug-2018

Revision: 3.4


Copyright: 2017, 2018
Page 2 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

Module List
Module identification: The module identifier (e.g. G7117A) can be found on the lower right side of
the module front cover.
Module

Instrument Description

G1328B

Manual Injector

G1364F

1260 Infinity II Analytical Fraction Collector

G5628A

Bio-inert Manual Injector

G5654A


Bio-inert Quaternary Pump

G5664B

1260 Infinity II Bio-inert Fraction Collector

G5668A

Bio-inert Multisampler

G7102A

Evaporative Light Scattering Detector

G7104A/C

Flexible Pump

G7110B

Isocratic Pump

G7111A/B

Quaternary Pump VL / Quaternary Pump

G7112B

Binary Pump


G7114A/B

Variable Wavelength Detector

G7115A

Diode Array Detector WR

G7116A/B

Multicolumn Thermostat

G7117A/B/C

Diode Array Detector FS / Diode Array Detector / Diode Array Detector HS

G7120A

High Speed Pump

G7121A/B

Fluorescence Detector / Fluorescence Detector Spectra

G7122A

Degasser

G7129A/B/C


Vialsampler VL/Vialsampler

G7162A/B

Refractive Index Detector

G7165A

Multiple Wavelength Detector

G7167A/B

Multisampler

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 3 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7104A/C
G7104A/C Procedure for 1290 Infinity II Easy Maintenance and LongLife Pump Heads
This procedure requires a torque wrench.

 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the pump head assembly.
 Follow the steps provided in the 1290 Infinity II Easy Maintenance Pump Head TechNote and
the 1290 Infinity II Long Life Pump Head TechNote.
 Replace the pump seals.
 Replace the seal wash seals.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Connect the solvent lines.
 Purge the pump with isopropanol.
 Replace the high pressure filter frit.
 Replace the inline filter frit, if installed.
 Replace the seal wash pump cartridge.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.

G7104A Pump Head Exchange Procedure
 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the classic pump head assembly.
 Install the Infinity II Easy Maintenance pump head assembly.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Replace the high pressure filter frit.
 Replace the inline filter frit, if installed.
 Replace the seal wash pump cartridge.
 Connect the solvent lines.

 Purge the pump with isopropanol.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.
Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 4 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist
G7104A Alternative Procedure (Classic Pump Heads)
This procedure requires the 1290 Infinity Pump head service kit.
 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the pump head assembly.
 Follow the steps provided in the Agilent 1290 Infinity Pump Head Maintenance Technote
 Replace the pump seals.
 Replace the seal wash seals and gasket.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Connect the solvent lines.
 Purge the pump with isopropanol.
 Replace the high pressure filter frit.

 Replace the inline filter frit, if installed.
 Replace the seal wash pump cartridge.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 5 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7110B, G7111A/B, G5654A
 Section NOT Applicable
 Power off the pump.
 Remove pump head.
 Disassemble the pump head.
 Remove and clean pistons.
 Clean support rings.
 Replace piston seals.
 Replace wash seals and gaskets, if the seal wash option is installed.
 Reassemble the pump head and install it.

 Reconnect the AIV to the connector (if installed).
 Replace the seal wash peristaltic pump if active seal wash is installed.
 Power on the module.
 For standard (PTFE) or PE seals perform the seal wear-in procedure.
 While running, begin maintenance on the sampler and column compartment if applicable.
 Purge valve: replace the PTFE frit, replace the seal cap assembly.
 If seal wash option is installed: Review the correct installation of seal wash function in





non-recycle mode. Manually activate the seal wash and check if there is a flow.
Perform the System Pressure Test (Agilent Lab Advisor software).
Perform the Leak Rate Test (Agilent Lab Advisor software).
Save or print-out results.
Reset EMF counters.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 6 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist


G7112B
 Section NOT Applicable
 Power off the pump.
 Remove pump heads.
 Disassemble the pump head.
 Remove and clean pistons.
 Clean support rings.
 Replace piston seals.
 Replace wash seals and gaskets, if the seal wash option is installed.
 Reassemble the pump heads and install them. Reconnect the Active Inlet Valves to the












connectors.
Replace the seal wash peristaltic pump if active seal wash is installed.
Power on the module.
For standard (PTFE) or PE seals perform the seal wear-in procedure.
While running, begin maintenance on the sampler and column compartment if applicable.
Purge valve: replace the PTFE frit, replace the seal cap assembly.
If seal wash option is installed: Review the correct installation of seal wash function in

non-recycle mode. Manually activate the seal wash and check if there is a flow.
Perform the system pressure test (Agilent Lab Advisor Software).
Perform a Pump Leak Rate Test (Agilent Lab Advisor software)
Perform a pump elasticity calibration (Agilent Lab Advisor Software).
Save or print-out results.
Reset EMF counters.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 7 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7120A
G7120A Procedure for 1290 Infinity II Easy Maintenance and LongLife Pump Heads
This procedure requires a torque wrench.
 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the pump head assembly.
 Follow the steps provided in the 1290 Infinity II Easy Maintenance Pump Head TechNote and
the 1290 Infinity II Long Life Pump Head TechNote.
 Replace the pump seals.

 Replace the seal wash seals.
 Replace the filter frit (PTFE frits (pack of 5) (01018-22707)) and seal cap (Seal cap
(5067-4728)) of the high pressure filter assembly.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Connect the solvent lines.
 Purge the pump with isopropanol.
 Replace the seal wash pump cartridge.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform the system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.

G7120A Pump Head Exchange Procedure
 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the classic pump head assembly.
 Install the Infinity II Easy Maintenance Pump Head assembly.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Connect the solvent lines.
 Purge the pump with isopropanol.
 Replace the seal wash pump cartridge.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform the system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.

Issued: 10-Aug-2018

Revision: 3.4


Copyright: 2017, 2018
Page 8 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist
G7120A Alternative Procedure (Classic Pump Heads)
This procedure requires the 1290 Infinity Pump head service kit
 Section NOT Applicable
 Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software).
 Remove or lift up solvent inlets in order to avoid solvent spilling.
 Remove the pump head assembly.
 Follow the steps provided in the Agilent 1290 Infinity Pump Head Maintenance Technote
 Replace the pump seals.
 Replace the seal wash seals and gasket.
 Replace the filter frit (PTFE frits (pack of 5) (01018-22707)) and seal cap (Seal cap
(5067-4728)) of the high pressure filter assembly.
 Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software).
 Connect the solvent lines.
 Purge the pump with isopropanol.
 Replace the seal wash pump cartridge.
 Perform pump leak rate test (Agilent Lab Advisor Software).
 Perform the system pressure test (Agilent Lab Advisor Software).
 Save or print-out results.
 Reset EMF counters.

Issued: 10-Aug-2018


Revision: 3.4

Copyright: 2017, 2018
Page 9 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7129A/B/C
 Section NOT Applicable
 Verify the proper function of the drawers.
 Replace the rotor seal.
 Replace the needle assembly and the high pressure needle seat assembly.
 Replace the peristaltic pump cartridge.
 Check the finger of the vial gripper.
 Verify the proper installation of the sample loop. Especially check if sample loop is moving













smoothly with the safety cover installed.
Check the proper function of the needle wash. Verify correct positioning of the drain tube.
Clean the instrument, especially the ventilation.
Clean the ICC heater (if applicable).
Clean the Sample Cooler, especially the ventilation and verify that the drain tube is positioned
correctly (if applicable).
Clean the Sample Thermostat, especially the ventilation and verify that the drain tube is
positioned correctly (if applicable).
After maintaining the multicolumn thermostat or the ICC heater, purge both modules with
isopropanol for 5 min.
Cap the outlet of the vialsampler with a blank nut ( Blank Nut SL (5067-6127)).
Perform a System Pressure Test.
Perform a System Pressure Test with ICC heater (if applicable).
Perform a Sample Cooler Test (if applicable).
Perform a ICC Heater Test (if applicable).

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 10 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist


G1328B/G5628A
 Section NOT Applicable
 Replace rotor seal.
 Replace the stator face assy (G1328C: valve 5067-4191, G5628A: valve 5067-4158).
 Purge injector and column compartment with isopropanol for 5 min.
 Cap the valve outlet with a blank nut.
 Perform a (System) Pressure Test. Take care for the maximum pressure of the valve and

installed loop. (PEEK loop max. pressure is 200 bar).

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 11 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7167A/B, G5668A
 Section NOT Applicable
 Verify the proper function of the drawer(s).
 Check the needle position of the Multisampler (see LadAdvisor: Instrument control > Special















Commands > Check Needle position). If the needle tip position is off the center perform an
auto referencing under operating conditions.
Replace the rotor seal(s).
Replace the needle assembly(ies) and the high pressure needle seat assembly(ies).
Verify the proper installation of the sample loops especially check if sample loops are kinked
or bended.
Replace the peristaltic pump cartridge (if applicable).
Check the proper function of the needle wash and seat backflushing (if applicable). Verify
correct positioning of the drain tube.
Clean the instrument, especially the ventilation.
Clean the Sample Cooler, especially the ventilation and verify that the drain tube is positioned
correctly (if applicable).
Clean the Sample Thermostat, especially the ventilation and verify that the drain tube is
positioned correctly (if applicable).
After maintaining the multicolumn thermostat, purge both modules with isopropanol for
5 min.
Cap the outlet of the multisampler with a blank nut ( Blank Nut SL (5067-6127)). If this blank
nut is not immediatly availabe perform the test with the TCC included.

Perform a System Pressure Test (for a dual needle system you have to check the left and the
right flow path).
Perform a Sample Cooler Test (if applicable).

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 12 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7116A/B
 Section NOT Applicable
 Before disconnecting the columns, please record the columns and their position as a reference






for re-installation.
If a valve is installed, replace the rotor seal (see parts section for details).
Purge LC system with HPLC grade water for 10 min.
Perform a System Pressure Test. Use a ZDV to install a blank nut onto the heat exchanger's

outlet capillary. If a valve is installed use the blanking nut to block one of the outlet ports. In
case of a column selector valve use a ZDV-union to mount the blanking nut on the outlet port
capillary with the valve switched to the bypass channel.
Reset EMF counter (Valve switches).

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 13 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7114A/B
The maintenance procedure requires NO consumables.
 Section NOT Applicable
 Inspect flow cell for leaks.
 Perform holmium oxide test.
 Perform Intensity Test.
 Perform Wavelength Verification Test.
 Perform Dark Current Test.
 Perform Filter Test.

Issued: 10-Aug-2018


Revision: 3.4

Copyright: 2017, 2018
Page 14 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7115A, G7165A
The maintenance procedure requires NO consumables.
 Section NOT Applicable
 Inspect flow cell for leaks.
 Perform Filter Test and attach results to checklist.
 Perform dark current test.
 Perform holmium oxide test.
 Perform Intensity Test.
 Perform Slit Test.
 Perform Wavelength Verification Test.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 15 of 39

Agilent Technologies



Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7121A/B
The maintenance procedure requires NO consumables.
 Section NOT Applicable
 Inspect the tubing from cell outlet to waste (must be a PTFE tubing without restriction).
 Inspect flow cell for leaks.
 Perform the lamp Intensity Test and compare profile (no given limits).
 Perform the wavelength verification test using the raman band.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 16 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7162A/B
 Section NOT Applicable
 Purge the reference and sample cells with water for 5 min.
 Check diode balance with pure water in both, the reference and the sample cell, instrument


stable as specified. Adjust optical balance if value >±0.2 .

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 17 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7117A/B/C
The maintenance procedure requires NO consumables (2x Position Cell Drawer Assembly
G4212-40007 (Replace only if worn))
 Section NOT Applicable
 Inspect flow cell for leaks.
 Check the condition of the positioner cell drawer assembly.
 Perform Wavelength Verification Test.
 Perform Intensity Test.
 Perform Dark Current Test.
 Perform Slit Test (not available in G7117A/C).

Issued: 10-Aug-2018

Revision: 3.4


Copyright: 2017, 2018
Page 18 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7102A
 Section NOT Applicable
 Check communication of Control Software with the instrument
 Mass Flow Controller - Check control of gas flow
 Lamp Assembly - Check change in baseline from standby to run.
 Photo Detector Module - Check change in baseline from standby to run
 Check autozero function
 Evaporator Heater Assembly - Check evaporator temperature increases/decreases
 Nebuliser Heater Assembly - Check nebuliser temperature increases/decreases
 Peltier Assembly (on G7102A cooled only) - Check for cooling to sub-ambient temperatures
 Clean optics block assembly if baseline noise greater than 3 LSD units
 If cleaning the Optics did not achieve the results then proceed with the cleaning of the prism






assembly.
Check for dust around cooling fan, and clean if necessary

Replace the following parts:
 Evaporator Cartridge
 Gas Inlet Plug Assembly
Clean the Nebuliser Assembly if cleaning prism and optics assembly, and replacing above
parts does not resolve baseline noise issues
Return the instrument to initial conditions.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 19 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G1364F, G5664B
 Section NOT Applicable
 Replace the Inlet / Waste tubing.
 Replace the Valve to Needle tubing.
 Perform a Delay Calibration.
 Perform a checkout method with the following settings:
Sample:

Delay Calibrant 5190-8223


Eluent A:

0.1 % TFA in LC grade water

Eluent B:

0.1 % TFA in ACN

Flow:

25 mL/min (5 mL/min on 9.4 mm column)

Gradient:

0.0 min

98 % A

2%B

0.3 min

98 % A

2%B

3.5 min

2%A


98 % B

4.2 min

2%A

98 % B

4.3 min

98 % A

2%B

5.5 min

98 % A

2%B

Runtime:

5.5 min

Injection volume:

50 µL (10 µL on 9.4 mm column)

Peak Detector:


UV or DAD signal A, UV 310 nm (+/- 4 nm, 5 Hz, no reference)

Peak Det. Mode:

Threshold (5 mAU) and Slope (50 mAU)

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 20 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

Guidance

If the PM service is performed prior to a qualification service, then use the qualification
procedure as a guide for final instrument set up and checkout.

Service Review
 Attach available reports or printouts to this documentation.
 Record the PM service activity in the customer’s instrument records/logbook
 Update/reset instrument maintenance counters as appropriate
 Affix the PM sticker to the system or instrument logbook based on the customer’s request.
 Complete the Service Engineer Comments section below if there are additional comments.

 Review the service and any test results with the customer.
 If the instrument firmware was updated, record the details of the change in the Service

Engineer Comments box below or if necessary, in the customer’s IQ records.

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 21 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

Pump Test Results Table
 Section NOT applicable
Test Description

Expected Test Result

System Pressure Test

Pass

Leak Rate Test (1260 Infinity II pumps)


Pass

Pump Leak Rate Test

Pass

Pump Self Test (if necessary)

Pass

Pump elasticity calibration (G7112B)

Pass

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 22 of 39

Actual Test Result

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7104A/C, G7120A Parts List Table

 Section NOT applicable
Part Description

Product/Model #
where used

Easy Maintenance Pump Head Channel
A

G4220A

Easy Maintenance Pump Head Channel
B

G4220A

Easy Maintenance Pump Head Quat

G4204A

Part Number

Quantity
Consumed

G4220-60300

G7120A
G4220-60310


G7120A
G4204-60300

G7104A
Long Life Pump Head Channel A

G4220A

G4220-60350

G7120A
Long Life Pump Head Channel B

G4220A

G4220-60360

G7120A
Long Life Pump Head Quat

G4204A

G4204-60350

G7104A/C
Wash Seal PE

1100/1200 Series Pumps

0905-1718


1200 Infinity Series
Pumps
InfinityLab LC Series
Pumps
all 1120 and 1220
systems
PE seals (pack of 2)

1100/1200 Series Pumps

0905-1420

1200 Infinity Series
Pumps
InfinityLab LC Series
Pumps
all 1120 and 1220
systems
Gasket, seal wash (pack of 6)

G7104A/C

5062-2484

G7120A
PTFE frits (pack of 5)

G7120A


01018-22707

Frit for pump outlet filter 2/pk

G7104A/C

5067-5716

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 23 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist
Part Description

Product/Model #
where used

Part Number

Frit 0.3 µm for inline filter, 5/pk

G7104A/C


5023-0271

Peristaltic pump with Pharmed tubing

1100/1200 Series Pumps

5065-4445

Quantity
Consumed

1200 Infinity Series
Pumps
InfinityLab LC Series
Pumps
1100/1200 Series
Autosamplers
1200 Infinity HiP
Samplers
InfinityLab LC Series
Samplers
PM Kit for LL/EM Heads (Quat. Pumps)

G7104A/C

G7104-68741

G4204A
PM Kit for LL/EM Heads (Binary

Pumps)

Issued: 10-Aug-2018

Revision: 3.4

G7120A

G7120-68741

G4220A

Copyright: 2017, 2018
Page 24 of 39

Agilent Technologies


Agilent InfinityLab LC Series
Preventive Maintenance Checklist

G7110B, G7111A/B, G7112B, G5654A Parts List Table
 Section NOT applicable
Part Description

Product/Model #
where used

Part Number


1260 Iso/Quat/1220 Pump PM kit
includes seal cap assemblies, standard
piston seals, film washers and PTFE
frits (use this kit for reversed phase
applications)

G7110B

1260 Binary Pump PM kit includes seal
cap assemblies, standard piston seals,
film washers and PTFE frits (use this kit
for reversed phase applications)

G7112B

G1312-68741

PM Kit for Bio-inert Quaternary Pump
includes bio-inert piston seals, seal cap
assemblies, film washers, bio-inert
wash seals, PTFE frits and the
peristaltic pump.

G5654A

G5611-68741

Seal Wash PM kit for 1260 Pumps
includes film washers and standard
wash seals.


G7110B

G1310-68742

Quantity
Consumed

G1310-68741

G7111A/B

G7111A/B
G7112B

Peristaltic pump with Pharmed tubing

1100/1200 Series Pumps

5065-4445

1200 Infinity Series Pumps
InfinityLab LC Series Pumps
1100/1200 Series
Autosamplers
1200 Infinity HiP Samplers
InfinityLab LC Series
Samplers
PE seals (pack of 2)


InfinityLab LC Series Pumps

0905-1420

Wash Seal PE

1100/1200 Series Pumps

0905-1718

1200 Infinity Series Pumps
InfinityLab LC Series Pumps
all 1120 and 1220 systems

Issued: 10-Aug-2018

Revision: 3.4

Copyright: 2017, 2018
Page 25 of 39

Agilent Technologies


×