Myocardial
Perfusion
Artifacts
Myocardial Perfusion Artifacts
Dark Rim Along Endocardial Border
From Systolic Motion
Reference: Storey, P. et al, MRM 48:1028-1036, 2002
Myocardial Perfusion Artifacts
Early
• Dark rim is
exaggerated
during systole
due to motion.
• Dark rim is
exaggerated
during early
maximal blush
because of
increased
susceptibility.
Late
Increasing Contrast Washout Time
Dark Rim Along Endocardial Border
Systole
Diastole
Increasing Cardiac Phases
Myocardial Perfusion Artifacts
Dark Rim Artifact in Phase-Encoding Direction
Reference: Univ Minn, Jerosh-Herold, M., et al
Increasing Contrast Washout Time
PE
PE
Early
PE
PE
PE
PE
PE
PE
All images at
TT = 255ms,
RR ~ 800ms
Late
•Phase-encoding direction typically has lower resolution than readout direction
•Lower resolution leads to stronger Gibbs ringing artifacts
•Lower resolution leads to more intravoxel dephasing due to susceptibility
Myocardial Perfusion Artifacts
Without Rotated FOV, iPAT Artifacts May Cause
False Perfusion Defects in Postero-Lateral Wall
Without Rotated FOV
With Rotated FOV
Myocardial Perfusion Artifacts
Inadquate RF Spoiling
appears as little dancing
dark lines in septal and
antero-lateral walls
To avoid RF Spoiling Artifacts:
• Use RF spoiling ON
• Use low Flip (10-12 deg)
Reference: Crawley, A.P. et al,
MRM Nov 8 (3) 248-260, 1988
Perfusion Quality Control
Is the bolus tight enough?
Good Bolus
• Strong bloodpool blush
• Quick bloodpool washout
• Discernable 2nd Pass
Bad Bolus:
• Leaky IV or low njection rate
• High cardiac output
• Pulmonary disease
• Intracardiac shunt
Perfusion Quality Control
Is the bolus tight enough?
Good Bolus:
• Mean Curve ROI in LV
• Peaks in 5-10 sec
• Recirculation
Bad Bolus:
• Mean Curve ROI in LV
• Peaks in 20-30 sec
• No recirculation