Agilent 1290 Infinity II
Evaporative Light
Scattering Detector

User Manual

Agilent Technologies


Notices
© Agilent Technologies, Inc. 2014

Warranty

No part of this manual may be reproduced
in any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.

The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any


information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.

Manual Part Number
G7102-90000

Edition
08/2014
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn

This product may be used as a component of an in vitro diagnostic system if the system is registered with
the appropriate authorities and complies with the relevant regulations.
Otherwise, it is intended only for general laboratory use.

receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.

Safety Notices


CAUTION
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated conditions are fully understood and
met.

Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation
or contract clause. Use, duplication or disclosure of Software is subject to Agilent
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will


WA R N I N G
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.

Agilent 1290 Infinity II ELSD User Manual


In This Book

In This Book

This manual contains information on:
• Agilent 1290 Infinity II ELSD (G7102A)

1 Introduction to the Detector
This chapter gives an introduction to the module, instrument overview and
internal connectors.

2 Site Requirements and Specifications
This chapter provides information on environmental requirements, physical
and performance specifications.

3 Installing the Module
This chapter gives information about the installation of your ELSD.


4 LAN Configuration
This chapter provides information on connecting the module to the control
software.

5 Using the Module
This chapter explains the operational parameters of the ELSD.

6 Optimizing Performance
This chapter gives hints on how to optimize the performance or use
additional devices.

7 Troubleshooting and Diagnostics
This chapter gives an overview about the troubleshooting and diagnostic
features.

Agilent 1290 Infinity II ELSD User Manual

3


In This Book

8 Error Information
This chapter describes the meaning of error messages, and provides
information on probable causes and suggested actions on how to recover
from error conditions.

9 Maintenance
This chapter describes the maintenance of the ELSD.


10 Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.

11 Appendix
This chapter provides addition information on safety, legal and web.

4

Agilent 1290 Infinity II ELSD User Manual


Contents

Contents

1 Introduction to the Detector
Introduction to the ELSD
System Overview 9

7

8

2 Site Requirements and Specifications

17

Pre-installation Requirements 18
Site Requirements 19

Physical Specifications 22
Performance Specifications 23
3 Installing the Module

25

Unpacking the Module 26
Installing the Module 28
4 LAN Configuration

33

What You Have to Do First 34
TCP/IP Parameter Configuration 35
PC and Agilent OpenLAB ChemStation and EZChrom Setup
5 Using the Module

49

Before Using the Detector 50
Instrument Controls 51
Operational Parameters 55
Controlling the ELSD During an Injection 59
Controlling the ELSD within ChemStation or EZChrom
General Considerations 63

Agilent 1290 Infinity II ELSD User Manual

43


62

5


Contents

6 Optimizing Performance

67

Do’s and Don’ts of ELS Detection 68
Location of the Detector Module 69
Pumping systems 70
Mobile phase priming 71
7 Troubleshooting and Diagnostics

73

Troubleshooting 74
Troubleshooting an HPLC System
General Problems 76

75

8 Error Information

83

What Are Error Messages 84

Module Specific Error Messages
9 Maintenance

85

95

Introduction to Maintenance 96
Cautions and Warnings 97
Cleaning the Module 98
Inspection of Cables 99
Drying the Diffuser 100
Cleaning the Nebuliser 101
Cleaning Evaporator Tube 102
Putting the Instrument into Storage
Updating Detector Firmware 104
10 Parts and Materials for Maintenance
Identifying Parts and Materials
11 Appendix

103

109

110

111

General Safety Information 112
Waste Electrical and Electronic Equipment Directive

Radio Interference 116
Agilent Technologies on Internet 117

6

115

Agilent 1290 Infinity II ELSD User Manual


Agilent 1290 Infinity II ELSD User Manual

1
Introduction to the Detector
Introduction to the ELSD

8

System Overview 9
Basic Principles of Operation 9
Operational Parameters 12
Overview of ELS Detector 13

This chapter gives an introduction to the module, instrument overview and
internal connectors.

Agilent Technologies

7



1

Introduction to the Detector
Introduction to the ELSD

Introduction to the ELSD

The Evaporative Light Scattering Detector is a unique and highly sensitive
detector for semi- volatile and non- volatile solutes in a liquid stream. It
is mainly used as a concentration detector for High Performance Liquid
Chromatography (HPLC). The solvent stream containing the solute material
is nebulized and carried by a gas flow through an evaporation chamber.
The solvent is volatilized, leaving a mist of solute particles that scatter
light to a photosensitive device. The signal is amplified and a voltage
output provides the concentration of the solute particles passing through
the light.
The Agilent 1290 Infinity II ELSD may be used alone, or as one of several
detectors in a GPC or HPLC system. As the solvent or eluent is
evaporated in the course of the analysis, the Agilent 1290 Infinity II ELSD
must be the last in series if used in conjunction with other detectors. If
the Agilent 1290 Infinity II ELSD is being used as the last detector in a
series, care must be taken not to exceed the recommended backpressure
in detector cells in other units.
This manual instructs the user in the installation and operation of the
Agilent 1290 Infinity II ELSD for stand- alone use and control using
Agilent OpenLAB (ChemStation and EZChrom editions) and Agilent
ChemStation.

8


Agilent 1290 Infinity II ELSD User Manual


Introduction to the Detector
System Overview

1

System Overview

Basic Principles of Operation
Nebulization
The eluent inlet is connected to the nebuliser via a short length of
stainless steel capillary tube. The incoming eluent stream passes through
the heated nebuliser and is mixed with the incoming nebuliser gas stream.
The mixed gas and eluent stream form an aerosol plume containing a
uniform dispersion of droplets that then passes as a continuous flow into
the evaporator section. Any larger droplets or the inefficiently nebulized
droplets collect in the nebuliser chamber waste trap and then drain off via
the waste outlet into a collection bottle.

Evaporation
After nebulization the atomized spray is propelled through the evaporation
tube assisted by the carrier gas. In the evaporator section the solvent is
removed leaving a stream of dry particles of the analyte. A diffuser
located in the evaporator assists in the drying of the particles, acting as
an efficient heat exchanger, prevents ballistic particles reaching the
scattering chamber and randomizes the particle plume. The ELS Detector
uses patented gas flow technology in the evaporation zone to aid

evaporation at low temperatures. By adding a stream of dry nitrogen
(evaporation gas) at the entrance of the evaporator tube less volatile
solvents (for example, water) are easily evaporated. This evaporation gas
is controlled by the user and facilitates sub- ambient operation.

Agilent 1290 Infinity II ELSD User Manual

9


1

Introduction to the Detector
System Overview

Detection
Light in the optical chamber is passed through the instrument at right
angles to the direction of particle flow. A light trap is located opposite the
source of light to capture the transmitted incident beam eliminating
internal reflections within the instrument body. When pure solvent is
being evaporated, only its vapor passes through the light path and the
amount of light scattered to the photomultiplier is small and gives a
constant baseline response. When a non- volatile solute is present a
particle cloud passes through the light path, causing light to be scattered.
This scattered light enters the optical aperture of the detection system and
generates a signal response from the photodiode in real time. The quantity
of light detected is dependent on the solute concentration and solute
particle size distribution.

&#CZWja^oVi^dc


Figure 1

10

'#:kVedgVi^dc

(#9ZiZXi^dc

Principles of Operation

Agilent 1290 Infinity II ELSD User Manual


1

Introduction to the Detector
System Overview

Theory
There are four main processes, by which the path of electromagnetic
radiation or light can change direction, when passing through a medium
containing a suspended particulate phase, see Figure 2 on page 11.

GZ[aZXi^dc

B^Z

Figure 2


GZ[gVXi^dc

GVnaZ^\]

ELSD Scattering Mechanisms

The importance of each of these processes depends on the radius of the
particle (r) compared to the wavelength (λ) of the incident light. Rayleigh
scattering is predominant when r/λ is < 5·10- 2. When particle dimensions
are greater than λ/20 they no longer behave as point sources, and Mie
scattering becomes predominant. Once particle size approaches the
wavelength of incident light then reflection and refraction begin to prevail.
The relative importance of refraction and reflection can be understood by
examining the effects of the incident light on a single spherical particle
whose equilateral axis lies in the same plane as the photodetector and
light source. With this configuration, refraction is of greater significance
than reflection. The majority of organic compounds have refractive indices
between 1.3 and 1.5. Changes in the refractive index within this range will
not greatly affect the quantity of light reaching the detector. This accounts
for similarities in the sensitivity of the instrument to various compounds.

Agilent 1290 Infinity II ELSD User Manual

11


1

Introduction to the Detector
System Overview


Operational Parameters
The ELS Detector responds to all compounds that are less volatile than
the mobile phase and is independent of a compound’s optical properties.
It therefore provides advantages over other spectroscopic detectors for
detecting compounds that are deficient in a UV chromophore or
fluorophore.
The removal of aqueous mobile phase within an ELSD is typically achieved
by setting the evaporator temperature to the eluent’s boiling point (for
example 100 °C) in order to remove the solvent. For non- volatile
compounds, operating at these high temperatures maximizes the signal
response.
However, at these temperatures volatile and semi- volatile compounds are
destroyed and are therefore not detected. This is problematic for small
molecules, such as pharmaceuticals and drug candidates.
The ELS detector is specifically designed to evaporate difficult solvents at
ambient and sub- temperatures in order to maximize detection of
semi- volatile compounds.
The ELSD has patented technology that reduces the evaporation time of
highly aqueous solvents at low temperature, and also prevents the
evaporation tube becoming saturated, which would otherwise prevent
further evaporation occurring.
Using this patented evaporation gas technology, a 20 µm droplet of water
at 30 °C can be dried ca. 3x faster than just temperature alone. Using the
Agilent 1290 Infinity II ELSD (G7102A), water can be evaporated as low
as 20 °C, providing maximum sensitivity to thermally sensitive
compounds. For maximum sensitivity of non- volatile compounds the
evaporation gas can be turned off at higher evaporation temperatures.
Therefore, unlike other ELS detectors, where the evaporator temperature
is set according to the type of mobile phase, the ELS detector evaporator

temperature is independent of the mobile phase. Consequently, the ELS
detector can be set at 30 °C for all types of mobile phase provided the
evaporator gas flow is adjusted accordingly. This method of operation
ensures that the ELSD sensitivity is maximized even for low molecular
weight compounds.
To prevent against unnecessary gas usage, a controlled gas shut- off valve
is integrated into the detector gas manifold. This will only allow gas to
pass into the instrument when in RUN mode. Should the instrument
default to STANDBY mode the gas will reduce to a default value of
1.2 SLM for 15 min before closing.
12

Agilent 1290 Infinity II ELSD User Manual


Introduction to the Detector
System Overview

1

Overview of ELS Detector

&

'

(

)


Figure 3

ELSD overview (front)

1

Front screen display

2

Keypad

3

Eluent inlet

4

Solvent waste outlet

Agilent 1290 Infinity II ELSD User Manual

13


1

Introduction to the Detector
System Overview


&

'
(
.

)
*

-
+
,

Figure 4

14

ELSD overview (rear)

1

Vapor Sensor vent

2

Serial RS232 port

3

LAN connector


4

Firmware button

5

I/O connector (Remote Start input)

6

Mains Switch

7

Mains Input

8

Gas Inlet port

9

Exhaust port

Agilent 1290 Infinity II ELSD User Manual


Introduction to the Detector
System Overview

Table 1

ELS Detector I/O connections

Inputs

Output

NOTE

1

I/O description

Pin number

Timetable Start

14 & ground

Injection Sync

13 & ground

Remote A/Z

7 & ground

Pump stop contact closure –
normally open


3 & 10

Ground (to case)

1, 5, 6, 11

In order to make appropriate remote start and A/Z connections from a third party LC, a third
party remote start cable for Dimension software (Remote start cable (PL0890-0350)) can be
purchased from Agilent Technologies.

Agilent 1290 Infinity II ELSD User Manual

15


1

16

Introduction to the Detector
System Overview

Agilent 1290 Infinity II ELSD User Manual


Agilent 1290 Infinity II ELSD User Manual

2
Site Requirements and Specifications

Pre-installation Requirements
Site Requirements

18

19

Physical Specifications

22

Performance Specifications 23
Specifications 23
Specification Conditions 24

This chapter provides information on environmental requirements, physical and
performance specifications.

Agilent Technologies

17


2

Site Requirements and Specifications
Pre-installation Requirements

Pre-installation Requirements


For a detailed description of the environmental and operating
requirements of the ELSD, please refer to the Site Preparation Checklist.
This manual will give you an overview of the ELS detector, and describe
its operation in more detail.

18

Agilent 1290 Infinity II ELSD User Manual


Site Requirements and Specifications
Site Requirements

2

Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument, please refer to the Site Preparation Checklist for more details.

Laser Safety
The Agilent 1290 Infinity II ELSD (G7102A) is classified as a "Laser Class
1" product (IEC825- 1, CFR1040.10 & 1040.11). During normal operation of
the ELSD no laser light is accessible to the user.

WA R N I N G

Eye damage by laser light
Hazardous laser light can injure eyes.
➔ Do not remove covers and interlocks.
➔ Observe and note the laser warning signs carefully.


Power Considerations
Check the operating voltage of your instrument on the IEC inlet fuse
holder on rear of unit.

WA R N I N G

Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔ Connect your instrument to the specified line voltage only.

CAUTION

Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
➔ Make sure the power connector of the instrument can be easily reached and

unplugged.
➔ Provide sufficient space behind the power socket of the instrument to unplug the

cable.

Agilent 1290 Infinity II ELSD User Manual

19


2


Site Requirements and Specifications
Site Requirements

Power Cords
Your detector is delivered with a power cord which matches the wall
socket of your particular country or region. The plug on the power cord
which connects to the rear of the instrument is identical for all types of
power cord.

WA R N I N G

Absence of ground connection or use of unspecified power cord
The absence of ground connection or the use of unspecified power cord can lead to
electric shock or short circuit.
➔ Never operate your instrumentation from a power outlet that has no ground

connection.
➔ Never use a power cord other than the Agilent Technologies power cord designed

for your region.

WA R N I N G

Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
➔ Never use cables other than the ones supplied by Agilent Technologies to ensure

proper functionality and compliance with safety or EMC regulations.


WA R N I N G

Unintended use of supplied power cords
Using power cords for unintended purposes can lead to personal injury or damage of
electronic equipment.
➔ Never use the power cords that Agilent Technologies supplies with this instrument

for any other equipment.

20

Agilent 1290 Infinity II ELSD User Manual


Site Requirements and Specifications
Site Requirements

2

Bench Space
The module dimensions and weight (see Table 2 on page 22) allow you to
place the module on almost any desk or laboratory bench. It needs an
additional 25 cm (10 inches) of space on either side and approximately
15 cm (5.9 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the
bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.

Agilent 1290 Infinity II ELSD User Manual


21


2

Site Requirements and Specifications
Physical Specifications

Physical Specifications

Table 2

22

Physical Specifications

Type

Specification

Comments

Weight

11 kg (non-cooled), 13 kg (cooled)

Dimensions
(height × width × depth)

415 x 200 x 450 mm

(16.3 x 7.9 x 17.7 inches)

Line voltage

100 – 240 V~, ± 10 %

Line frequency

50 or 60 Hz, ± 5 %

Power consumption

150 W (max)

Ambient operating
temperature

10–35 °C (50–95 °F)

Ambient non-operating
temperature

-40 – 70 °C (-40 – 158 °F)

Humidity

< 80 % r.h. at 40 °C (104 °F)

Operating altitude


Up to 2000 m (6562 ft)

Non-operating altitude

Up to 4600 m (15091 ft)

For storing the module

Safety standards:
IEC, CSA, UL

Installation category II, Pollution degree 2

For indoor use only.

Wide-ranging
capability

Maximum

Non-condensing

Agilent 1290 Infinity II ELSD User Manual


Site Requirements and Specifications
Performance Specifications

2


Performance Specifications

Specifications
The instrument is suitable for indoor use only and is classified suitable
under the following categories (EN 61010- 1):2010
• Installation category II
• Pollution degree 2
• Safety class 1
Table 3

Performance Specification of the ELS Detector

Type

Specification

Light Source
G7102A

LASER 405 nm, 10 mW (Class 3B)

Detector

Dual PMT with digital signal processing

Nebuliser

OFF, 25 – 90 °C

Evaporator

Non-cooled

OFF, 25 – 120 °C

Cooled

OFF, 10 – 80 °C

Gas Flow Range

0.9 – 3.25 SLM (controlled gas shut-off)

Short Term Noise

<0.1 mV under specified condtions. See “Specification
Conditions” on page 24.

Drift

<1 mV/h under specified condtions. See “Specification
Conditions” on page 24.

Operating Pressure

60 – 100 psi (4.1 – 6.9 bar)

Eluent Flow range

0.2 – 5.0 mL/min


Digital Output

10, 40 or 80 Hz (24 bit)

Agilent 1290 Infinity II ELSD User Manual

23


2

Site Requirements and Specifications
Performance Specifications
Table 3

Performance Specification of the ELS Detector

Type

Specification

Communication

Ethernet
Serial (RS232)
Remote Start Input
Pump Stop: 1 Contact closure

PC control (software)


ELSD driver for OpenLAB ChemStation edition
ELSD driver for OpenLAB EZChrom edition

Remote operation

Remote Start Input

Safety features

Gas shut-off Valve, Leak Detection, Laser Interlock

Specification Conditions
ASTM: “Evaporative Light Scattering Detectors Used in Liquid
Chromatography”.
Table 4

Reference conditions

Gas flow

1.6 SLM

Neb temperature

40 °C

Evaporator temperature

40 °C


Data Rate

40 Hz

Smoothing

1

ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h)
over one hour period. Our published drift specification is based on these
conditions. Larger ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency
and the amplitude of the temperature changes to below 1 °C/h (1.8 °F/h).
Turbulences around one minute or less can be ignored. ASTM
measurements require that the detector should be turned on enough time
before start of testing.

24

Agilent 1290 Infinity II ELSD User Manual


Agilent 1290 Infinity II ELSD User Manual

3
Installing the Module
Unpacking the Module 26
Delivery Checklist 27
Installing the Module 28

Connecting the Detector to your LC System 28
Checking Operation of the ELS Detector 31

This chapter gives information about the installation of your ELSD.

Agilent Technologies

25


×