Tải bản đầy đủ (.pdf) (31 trang)

Rotational and Vibrational Spectroscopy pdf

Bạn đang xem bản rút gọn của tài liệu. Xem và tải ngay bản đầy đủ của tài liệu tại đây (2.12 MB, 31 trang )

1
Lecture Date: January 30
th
, 2008
Rotational and Vibrational
Spectroscopy
Vibrational and Rotational Spectroscopy
 Core techniques:
– Infrared (IR) spectroscopy
– Raman spectroscopy
– Microwave spectroscopy
2
The Electromagnetic Spectrum
 The basic!
 Microwave
 Infrared (IR)
The History of Infrared and Raman Spectroscopy
 Infrared (IR) Spectroscopy:
– First real IR spectra measured by Abney and Festing in 1880’s
– Technique made into a routine analytical method between 1903-
1940 (especially by Coblentz at the US NBS)
– IR spectroscopy through most of the 20
th
century is done with
dispersive (grating) instruments, i.e. monochromators
– Fourier Transform (FT) IR instruments become common in the
1980’s, led to a great increase in sensitivity and resolution
 Raman Spectroscopy:
– In 1928, C. V. Raman discovers that small changes occur the
frequency of a small portion of the light scattered by molecules.
The changes reflect the vibrational properties of the molecule


– In the 1970’s, lasers made Raman much more practical. Near-
IR lasers (1990’s) allowed for avoidance of fluorescence in
many samples.
W. Abney, E. R. Festing, Phil. Trans. Roy. Soc. London, 1882, 172, 887-918.
3
Infrared Spectral Regions
 IR regions are traditionally sub-divided as follows:
Region Wavelength
(), m
Wavenumber
(), cm
-1
Frequency
(), Hz
Near 0.78 to 2.5 12800 to 4000 3.8 x 10
14
to
1.2 x 10
14
Mid 2.5 to 50 4000 to 200 1.2 x 10
14
to
6.0 x 10
12
Far 50 to 1000 200 to 10 6.0 x 10
12
to
3.0 x 10
11
After Table 16-1 of Skoog, et al. (Chapter 16)

What is a Wavenumber?
 Wavenumbers (denoted cm
-1
) are a measure of frequency
– For an easy way to remember, think “waves per centimeter”
 Relationship of wavenumbers to the usual frequency and
wavelength scales:
Image from www.asu.edu


10000
1


cm
 Converting
wavelength () to
wavenumbers:
4
Rotational and Vibrational Spectroscopy: Theory
 Overview:
– Separation of vibrational and rotational contributions to energy is
commonplace and is acceptable
– Separation of electronic and rovibrational interactions
 Basic theoretical approaches:
– Harmonic oscillator for vibration
– Rigid rotor for rotation
 Terminology:
– Reduced mass (a.k.a. effective mass):
See E. B. Wilson, Jr., J. C. Decius, and P. C. Cross, “Molecular Vibrations”, Dover, 1955.

21
21
mm
mm



Rotational Spectroscopy: Theory
 Rotational energy levels can be
described as follows:
R. Woods and G. Henderson, “FTIR Rotational Spectroscopy”, J. Chem. Educ., 64, 921-924 (1987)
DJBJJ
3
)1()1()( 

crhB
2
0
2
8/


23
/4
c
BD


Where:
c is the speed of light

k is the Hooke’s law force constant
r
0
is the vibrationally-averaged bond length
The rotational constant:
The centrifugal distortion coefficient:
u
k
c
c


2
1

Example for HCl:
B
0
= 10.4398 cm
-1
D
0
= 0.0005319 cm
-1
r
0
= 1.2887 Å
 is the reduced mass
h is Planck’s constant


0
= 2990.946 cm
-1
(from IR)
k = 5.12436 x 10
5
dyne/cm
-1
For J = 0, 1, 2, 3…
5
Vibrational Spectroscopy: Theory
 Harmonic oscillator – based on the classical “spring”


m
hvE

2
1


m
is the natural frequency of the oscillator (a.k.a. the fundamental vibrational wavenumber)
k is the Hooke’s law force constant (now for the chemical bond)
u
k
m


2

1

v is the vibrational quantum number
h is Planck’s constant
 Since v must be a whole number (see Ex. 16-1, pg. 386):
 The potential energy function is:
2
2
1
)()(
eHO
rrkrE 
Note – all E are
potential energies (V)!
or
22
2
1
)()2()(
emHO
rrcrE 



kh
hE
m
2




k
12
103.5


and
(wavenumbers)
r is the distance (bond distance)
r
e
is the equilibrium distance
Vibrational Spectroscopy: Theory
 Potential energy of a harmonic oscillator:
Figure from Skoog et al.
6
Anharmonic Corrections
 Anharmonic motion: when the restoring force is not
proportional to the displacement.
– More accurately given by the Morse potential function than by the
harmonic oscillator equation.
– Primarily caused by Coulombic (electrostatic) repulsion as atoms
approach
 Effects: at higher quantum numbers, E gets smaller, and
the ( = +/-1) selection rule can be broken
– Double ( = +/-2), triple ( = +/-3), and higher order transitions
can occur, leading to overtone bands at higher frequencies (NIR)
2
)(
)1()(

e
rra
eMorse
ehcDrE


D
e
is the dissociation energy
e
m
hcD
c
a
2
)2(
2


Vibrational Coupling
 Vibrations in a molecule may couple – changing each
other’s frequency.
– In stretching vibrations, the strongest coupling occurs between
vibrational groups sharing an atom
– In bending vibrations, the strongest coupling occurs between
groups sharing a common bond
– Coupling between stretching and bending modes can occur when
the stretching bond is part of the bending atom sequence.
– Interactions are strongest when the vibrations have similar
frequencies (energies)

– Strong coupling can only occur between vibrations with the same
symmetry (i.e. between two carbonyl vibrations)
7
Vibrational Modes and IR Absorption
 Number of modes:
– Linear: 3n – 5 modes
– Non-linear: 3n – 6 modes
 Types of vibrations:
– Stretching
– Bending
 Examples:
– CO
2
has 3 x 3 – 5 = 4 normal
modes
Symmetric
No change in dipole
IR-inactive
Asymmetric
Change in dipole
IR-active
Scissoring
Change in dipole
IR-active
 IR-active modes require dipole changes during rotations
and vibrations!
Vibrational Modes: Examples
 IR-activity requires
dipole changes
during vibrations!

 For example, this
is Problem 16-3
from Skoog:
Inactive
Active
Active
Active
Inactive
Inactive
Active
8
IR Spectra: Formaldehyde
 Certain types of vibrations have distinct IR frequencies – hence the
chemical usefulness of the spectra
 The gas-phase IR spectrum of formaldehyde:
Formaldehyde spectrum from: />Results generated using B3LYP//6-31G(d) in Gaussian 03W.
 Tables and simulation results can help assign the vibrations!
(wavenumbers, cm
-1
)
Rayleigh and Raman Scattering
 Only objects whose dimension is ~1-1.5  will scatter EM
radiation.
 Rayleigh scattering:
– occurs when incident EM radiation induces an oscillating dipole in
a molecule, which is re-radiated at the same frequency
 Raman scattering:
– occurs when monochromatic light is scattered by a molecule, and
the scattered light has been weakly modulated by the
characteristic frequencies of the molecule

 Raman spectroscopy measures the difference between
the wavelengths of the incident radiation and the
scattered radiation.
9
The Raman Effect
 Polarization changes
are necessary to form
the virtual state and
hence the Raman
effect
 This figure depicts
“normal” (spontaneous)
Raman effects
H. A. Strobel and W. R. Heineman, Chemical Instrumentation: A Systematic Approach, 3
rd
Ed. Wiley: 1989.
hv
1
Scattering timescale ~10
-14
sec
(fluorescence ~10
-8
sec)
Virtual state
Virtual state
hv
1
Ground state
(vibrational)

 The incident radiation excites “virtual states” (distorted
or polarized states) that persist for the short timescale of
the scattering process.
Excited state
(vibrational)
hv
1
– hv
2
Stokes line
hv
1
– hv
2
Anti-Stokes line
More on Raman Processes
 The Raman process: inelastic scattering of a photon
when it is incident on the electrons in a molecule
– When inelastically-scattered, the photon loses some of its energy
to the molecule (Stokes process). It can then be experimentally
detected as a lower-energy scattered photon
– The photon can also gain energy from the molecule (anti-Stokes
process)
 Raman selection rules are based on the polarizability of
the molecule
 Polarizability: the “deformability” of a bond or a molecule
in response to an applied electric field. Closely related to
the concept of “hardness” in acid/base chemistry.
P. W. Atkins and R. S. Friedman, Molecular Quantum Mechanics, 3
rd

Ed. Oxford: 1997.
10
More on Raman Processes
 Consider the time variation of the dipole moment induced
by incident radiation (an EM field):
)()()( ttt




P. W. Atkins and R. S. Friedman, Molecular Quantum Mechanics, 3
rd
Ed. Oxford: 1997.
EM fieldInduced dipole moment
 Expanding this product yields:


tttt )cos()cos(cos)(
intint0
4
1
0















Rayleigh line
Anti-Stokes line Stokes line
polarizability
 If the incident radiation has frequency  and the
polarizability of the molecule changes between 
min
and

max
at a frequency 
int
as a result of this rotation/vibration:


ttt






coscos)(
0int
2

1



mean polarizability
 = 
max
- 
min
The Raman Spectrum of CCl
4
Figure is redrawn from D. P. Strommen and K. Nakamoto, Amer. Lab., 1981,43 (10), 72.
Observed in
“typical”
Raman
experiments

0
= 20492 cm
-1

0
= 488.0 nm
Anti-Stokes lines
(inelastic scattering)
-218
Raman shift cm
-1

0

= (
s
- 
0
)
-200
Stokes lines
(inelastic scattering)
-400400 200
218
314
-314
-459
459
0
Rayleigh line
(elastic scattering)
11
Raman-Active Vibrational Modes
 Modes that are more polarizable are more Raman-active
 Examples:
– N
2
(dinitrogen) symmetric stretch
 cause no change in dipole (IR-inactive)
 cause a change in the polarizability of the bond – as the bond gets
longer it is more easily deformed (Raman-active)
– CO
2
asymmetric stretch

 cause a change in dipole (IR-active)
 Polarizability change of one C=O bond lengthening is cancelled by
the shortening of the other – no net polarizability (Raman-inactive)
 Some modes may be both IR and Raman-active, others
may be one or the other!
The Raman Depolarization Ratio
 Raman spectra are excited by linearly polarized radiation
(laser).
 The scattered radiation is polarized differently depending
on the active vibration.
 Using a polarizer to capture the two components leads to
the depolarization ratio p:



I
I
p
 The depolarization ratio p can be useful in interpreting the
actual vibration responsible for a Raman signal.
12
Instrumentation for Vibrational Spectroscopy
 Absorption vs. Emission for IR spectroscopy:
– Emission is seldom used for chemical analysis
– The sample must be heated to a temperature much greater than its
surroundings (destroying molecules)
– IR emission is widely used in astronomy and in space applications.
 Two IR Absorption methods:
– Dispersive methods: Scanning of wavelengths using a grating
(common examples are double-beam, like a spectrometer

discussed in the optical electronic spectroscopy lecture).
– Fourier-transform methods: based on interferometry, a method of
interfering and modulating IR radiation to encode it as a function
of its frequency.
Radiation
Source
Sample
Wavelength
Selector
Detector
(transducer)
Radiation
Source
Interferometer Sample
Detector
(transducer)
Why Build Instruments for Fourier Transform Work?
 Advantages:
– The Jacqinot (throughput) advantage: FT instruments have
few slits, or other sources of beam attenuation
– Resolution/wavelength accuracy (Connes advantage):
achieved by a colinear laser of known frequency
– Fellgett (multiplex) advantage: all frequencies detected at
once, signal averaging
 These advantages are critical for IR spectroscopy
 The need for FT instruments is rooted in the detector
– There are no transducers that can acquire time-varying signals
in the 10
12
to 10

15
Hz range – they are not fast enough!
 Why are FT instruments not used in UV-Vis?
– The multiplex disadvantage (shot noise) adversely affects
signal averaging – it is better to multiplex with array detectors
(such as the CCD in ICP-OES)
– In some cases, technical challenges to building interferometers
with tiny mirror movements
13
Inteferometers for FT-IR and FT-Raman
 The Michelson
interferometer, the
product of a famous
physics experiment:
 Produces
interference
patterns from
monochromatic
and white light
Figures from Wikipedia.org
Inteferometers
 For monochromatic
radiation, the
interferogram looks like
a cosine curve
 For polychromatic
radiation, each
frequency is encoded
with a much slower
amplitude modulation

 The relationship
between frequencies:
 Example: mirror rate = 0.3 cm/s modulates 1000 cm
-1
light at 600 Hz
 Example: mirror rate = 0.2 cm/s modulates 700 nm light at 5700 Hz

c
v
f
M
2

Where:
 is the frequency of the radiation
c is the speed of light in cm/s
v
m
is the mirror velocity in cm/s
14
The Basics of the Fourier Transform
 The conversion from time- to frequency domain:
50 100 150 200 250
-1
-0.5
0.5
1
50
100
150

200
250
0.5
1
1.5
2
FT
50 100 150 200 250
-1.5
-1
-0.5
0.5
1
1.5
2
50
100
150
200
250
0.5
1
1.5
2
2.5






1
0
/2
1
N
k
Nikn
kn
ed
N
f




b
a
dtftKg

)(),()(
1
)texp(),( iωtK



Continuous:
Discrete:
FT
FTIR Spectrometer Design
Michelson

Interferometer
IR Source
Sample
Moving MirrorFixed Mirror
Beamsplitter
Detector
Interferogram
Fourier Transform - IR Spectrum
 It is possible to build a detector that detects multiple
frequencies for some EM radiation (ex. ICP-OES with CCD,
UV-Vis DAD)
 FTIR spectrometers are designed around the Michelson
interferometer, which modulates each IR individual
frequency with an additional unique frequency:
15
IR Sampling Methods: Absorbance Methods
 Salt plates (NaCl): for liquids (a drop) and small amounts of solids.
Sample is held between two plates or is squeezed onto a single plate.
 KBr/CsI pellet: a dilute (~1%) amount of sample in the halide matrix
is pressed at >10000 psi to form a transparent disk.
– Disadvantages: dilution required, can cause changes in sample
 Mulls: Solid dispersion of sample in a heavy oil (Nujol)
– Disadvantages: big interferences
 Cells: For liquids or dissolved samples. Includes internal reflectance
cells (CIRCLE cells)
 Photoacoustic (discussed later)
IR Sampling Methods: Reflectance Methods
 Specular reflection: direct
reflection off of a flat surface.
– Grazing angles

 Attenuated total reflection
(ATR): Beam passed through
an IR-transparent material with
a high refractive index, causing
internal reflections. Depth is
~2 um (several wavelengths)
 Diffuse reflection (DRIFTS): a
technique that collects IR
radiation scattered off of fine
particles and powders. Used
for both surface and bulk
studies.
Figures from />ATR
DRIFTS
16
IR Sources
 Nernst glower: a rod or cylinder made from several grams
of rare earth oxides, heated to 1200-2200K by an electric
current.
 Globar: similar to the Nernst glower but made from silicon
carbide, electrically heated. Better performance at lower
frequencies.
 Incandescent Wires: nichrome or rhodium, low intensity
 Mercury Arc: high-pressure mercury vapor tube, electric
arc forms a plasma. Used for far-IR
 Tungsten filament: used for near-IR
 CO
2
Lasers (line source): high-intensity, tunable, used for
quantitation of specific analytes.

IR Detectors
 Thermal transducers
– Response depends upon heating effects of IR radiation
(temperature change is measured)
 Slow response times, typically used for dispersive instruments or
special applications
 Pyroelectric transducers
– Pyroelectric: insulators (dielectrics) which retain a strong electric
polarization after removal of an electric field, while they stay
below their Curie temperature.
– DTGS (deuterated triglycine sulfate): Curie point ~47°C
 Fast response time, useful for interferometry (FTIR)
 Photoconducting transducers
– Photoconductor: absorption of radiation decreases electrical
resistance. Cooled to LN
2
temperatures (77K) to reduce thermal
noise.
– Mid-IR: Mercury cadmium telluride (MCT)
– Near-IR: Lead sulfide (NIR)
17
Raman Spectrometers
 The basic design dispersive Raman scattering system:
 Special considerations:
– Sources: lasers are generally the only source strong enough to
scatter lots of light and lead to detectable Raman scattering
– Avoiding fluorescence: He-Cd (441.6 nm), Ar ion (488.0 nm,
514.5 nm), He-Ne (632.8), Diode (782 or 830), Nd/YAG (1064)
Sample
Wavelength

Selector
Detector
(photoelectric transducer)
Radiation
source
(90° angle)
Modern Raman Spectrometers
 FT-Raman spectrometers – also make use of Michelson
interferometers
– Use IR (1 m) lasers, almost no problem with fluorescence for
organic molecules
– Have many of the same advantages of FT-IR over dispersive
– But, there is much debate about the role of “shot noise” and
whether signal averaging is really effective
 CCD-Raman spectrometers – dispersive spectrometers
that use a CCD detector (like the ICP-OES system
described in the Optical Electronic lecture)
– Raman is detected at optical frequencies!
– Generally more sensitive, used for microscopy
– Usually more susceptible to fluorescence, also more complex
 Detectors - GaAs photomultiplier tubes, diode arrays, in
addition to the above.
18
More on Raman
 Raman can be used to study aqueous-phase samples
– IR is normally obscured by H
2
O modes, these happen to be less
intense in Raman
– However, the water can absorb the scattered Raman light and

will damp the spectrum, and lower its sensitivity
 Raman has several problems:
– Susceptible to fluorescence, choice of laser important
– When used to analyze samples at temperatures greater than
250C, suffers from black-body radiation interference (so does
IR)
– When applied to darkly-colored samples (e.g. black), the Raman
laser will heat the sample, can cause decomposition and/or
more black-body radiation
Applications of Raman Spectroscopy
 Biochemistry: water is not strongly detected in Raman
experiments, so aqueous systems can be studied.
Sensitive to e.g. protein conformation.
 Inorganic chemistry: also often aqueous systems.
Raman also can study lower wavenumbers without
interferences.
 Other unique examples:
– Resonance Raman spectroscopy: strong enhancement (10
2

10
6
times) of Raman lines by using an excitation frequency close
to an electronic transition (Can detect umol or nmol of analytes).
– Surface-enhanced Raman (SERS): an enhancement obtained
for samples adsorbed on colloidal metal particles.
– Coherent anti-Stokes Raman (CARS): a non-linear technique
using two lasers to observe third-order Raman scattering – used
for studies of gaseous systems like flames since it avoids both
fluorescence and luminescence issues.

19
Applications of Raman Spectroscopy
 Raman in catalysis research (see C&E News, Oct. 13,
2006, pg. 59):
– Useful for the study of zeolite interiors
– Fluorescence can be a problem, but one approach is to use UV
light (257 nm) which avoids it just like switching to the IR (but at
the risk of decomposition) – See work from the Stair group at
Northwestern
– For uses of SERS: Catal. Commun 3 547 (2002).
 Raman microscopy: offers sub-micrometer lateral
resolution combined with depth-profiling (when combined
with confocal microscopy)
Comparison of IR and Raman Spectroscopy
 Advantages of Raman over IR:
– Avoids many interferences from solvents, cells and sample
preparation methods
– Better selectivity, peaks tend to be narrow
– Depolarization studies possible, enhanced effects in some cases
– Can detect IR-inactive vibrational modes
 Advantages of IR over Raman:
– Raman can suffer from laser-induced fluorescence and
degradation
– Raman lines are weaker, the Rayleigh line is also present
– Raman instruments are generally more costly
– Spectra are spread over many um in the IR but are compressed
into several nm (20-50 nm) in the Raman
 Final conclusion – they are complementary techniques!
20
Interpretation of IR and Raman Spectra

 General Features:
– Stretching frequencies are greater (higher wavenumbers) than
corresponding bending frequencies
 It is easier to bend a bond than to stretch it
– Bonds to hydrogen have higher stretching frequencies than those
to heavier atoms.
 Hydrogen is a much lighter element
– Triple bonds have higher stretching frequencies than double
bonds, which have higher frequencies than single bonds
 Strong IR bands often correspond to weak Raman bands
and vice-versa
Interpretation of IR and Raman Spectra
Characteristic Vibrational Frequencies for Common Functional Groups
Frequency (cm
-1
) Functional Group Comments
3200-3500 alcohols (O-H)
amine, amide (N-H)
alkynes (CC-H)
Broad
Variable
Sharp
3000 alkane (C-C-H)
alkene (C=C-H)
2100-2300 alkyne (CC-H)
nitrile (CN-H)
1690-1760 carbonyl (C=O) ketones, aldehydes,
acids
1660 alkene (C=C)
imine (C=N)

amide (C=O)
Conjugation lowers
amide frequency
1500-1570
1300-1370
nitro (NO
2
)
1050-1300 alcohols, ethers, esters,
acids (C-O)
See also Table 17-2 of Skoog, et al.
More detailed lists are widely available. See R. M. Silverstein and F. X. Webster, “Spectrometric Identification of Organic Compounds”, 6
th
Ed., Wiley, 1998.
21
IR and Raman Spectra of an Organic Compound
The IR and Raman spectra of
flufenamic acid (an analgesic/anti-
inflammatory drug):
CF
3
O OH
FT-IR Flufenamic acid Aldrich as recd
0.05
0.10
0.15
0.20
0.25
0.30
Abs

FT-Raman Flufenamic acid Aldrich as recd
0
10
20
30
40
50
60
Int
500 1000 1500 2000 2500 3000 3500
Raman shift (cm-1)
IR and Raman Spectra of an Organic Compound
The IR and Raman spectra of
flufenamic acid (an analgesic/anti-
inflammatory drug):
CF
3
O OH
FT-IR Flufenamic acid Aldrich as recd
0.05
0.10
0.15
0.20
0.25
0.30
Abs
FT-Raman Flufenamic acid Aldrich as recd
0
10
20

30
40
50
60
Int
200 400 600 800 1000 1200 1400 1600
Wavenumbers (cm-1)
Note – materials
usually limit IR
in this region
22
IR and Raman Spectra of an Organic Compound
The IR and Raman spectra of tranilast:
Tranilast Form I FV101031-171A1 FTIR
0.1
0.2
0.3
0.4
0.5
0.6
Abs
Tranilast Form I FV101031-171A1 FT-Raman
100
200
300
400
500
Int
500 1000 1500 2000 2500 3000 3500
Wavenumbers (cm-1)

O
O
N
H
O
OHO
C1
C6C2
C3
C4
C5
C7
N1
C8
C9
C10
C11
C12
C13
C14
C15
C16
C17
C18
H
3
C
H
3
C

O4
O5
O3
O2
O1
IR Frequencies and Hydrogen Bonding Effects
 IR frequencies are sensitive to
hydrogen-bonding strength and
geometry (plots of relationships
between crystallographic distances
and vibrational frequencies):
G. A. Jeffrey, An Introduction to Hydrogen Bonding, Oxford, 1997.
23
Applications of Far IR Spectroscopy
 Far IR is used to study low frequency vibrations, like those between
metals and ligands (for both inorganic and organometallic chemistry).
– Example: Metal halides have stretching and bending vibrations in the
650-100 cm
-1
range.
– Organic solids show “lattice vibrations” in this region
 Can be used to study crystal lattice energies and semiconductor
properties.
 The Far IR region also overlaps rotational bands, but these are
normally not detectable in condensed-phase work
Terahertz Spectroscopy
 A relatively new technique, addresses an unused portion
of the EM spectrum (the “terahertz gap”):
– 50 GHz (0.05 THz) to 3 THz (1.2 cm
-1

to 100 cm
-1
)
 Made possible with recent innovations in instrument
design, accesses a region of crystalline phonon bands
P. F. Taday and D. A. Newnham, Spectroscopy Europe, , www.spectroscopyeurope.com
G. Winnewisser, Vibrational Spectroscopy8 (1995) 241-253
24
Applications of Near IR Spectroscopy
 Near IR – heavily used in process chemistry
 Amenable to quantitative analysis usually in conjunction with
chemometrics (calibration requires many standards to be run)
 While not a qualitative technique, it can serve as a fast and useful
quantitative technique especially using diffuse reflectance
 Accuracy and precision in the ~2% range
 Examples:
– On-line reaction monitoring (food, agriculture, pharmaceuticals)
– Moisture and solvent measurement and monitoring
 Water overtone observed at 1940 nm
– Solid blending and solid-state issues
Near IR Spectroscopy
Figure from www.asdi.com. For more information see:
1. Ellis, J.W. (1928) “Molecular Absorption Spectra of Liquids Below 3 m”, Trans. Faraday Soc. 1928, 25, pp. 888-898.
2. Goddu, R.F and Delker, D.A. (1960) “Spectra-structure correlations for the Near-Infrared region.” Anal. Chem., vol. 32 no. 1, pp. 140-141.
3. Goddu, R.F. (1960) “Near-Infrared Spectrophotometry,” Advan. Anal. Chem. Instr. Vol. 1, pp. 347-424.
4. Kaye, W. (1954) “Near-infrared Spectroscopy; I. Spectral identification and analytical applications,” Spectrochimica Acta,vol. 6, pp. 257-287.
5. Weyer, L. and Lo, S C. (2002) Spectra-Structure Correlations in the Near-infrared, In Handbook of Vibrational Spectroscopy, Vol. 3, Wiley, U.K., pp. 1817-1837.
6. Workman, J. (2000) Handbook of Organic Compounds: NIR, IR, Raman, and UV-Vis Spectra Featuring Polymers and Surfactants, Vol. 1, Academic Press, pp. 77-197.
25
Near IR Spectrum of Acetone

 NIR taken in transmission mode (via a reflective gold plate) on a
Foss NIRsystems spectrometer
 Useful for quick solvent identification
Near IR Spectrum of Water (1
st
Derivative)
 1
st
derivative (and 2
nd
derivative) allows for easier identification of
bands

×